EMERGENCY SWITCHES
Home > Radiation Protection and Quality Assurance > Equipment Use & Quality Assurance > Quality Control Procedures > Emergency Switches
Can we please get your advice on this one question?
Emergency switches are designed for safety purposes. They control both
the beam and the mechanical systems in the treatment room. They should always
be functional. All Linac operators should know the location of emergency switches
and how and when to activate them.
Locations of the
emergency switches in a linear accelerator room:
- On each side of the
stand
- On each side of the
treatment couch
- On each wall of the
bunker
- On the entry to the
maze
- On the machine console
in the control room
- On the power supply to
the linear accelerator (in the control room).
- Emergency switches in
a linear accelerator room are activated when:
- There is a power
failure.
- There is fire outbreak.
- The treatment beam
fails to terminate.
Equipment malfunction.
- The linear accelerator
produces external radiation that is higher than the pre-set level.
- There is someone in
the treatment room when the linear accelerator is active.
- There is a danger to the
patient and has to be removed out of the treatment room immediately, that is:
- If the patient has an
acute medical event.
- If there is a
dangerous substance within the bunker.
- If the gantry or
treatment head is about to collide with the patient.
Procedures to follow when
the above incidences happen:
Power failure
- Terminate the beam and
record the reading on the backup counter at the time of power failure.
- Remove the patient out
of the treatment room.
- Calculate the
remaining dosage (in monitor units) to complete the treatment by subtracting
the backup counter reading from the prescribed dose.
- Once the power is
back, perform the daily checkout procedure to verify proper operation of the
equipment.
- Return the patient to
the treatment room to resume treatment.
- A medical physicist should
verify the remaining dose to be delivered.
Fire outbreak
- Press the Emergency
button.
- Turn off the main
circuit breaker and call the emergency team.
- Remove the patient
from the treatment room.
- If the fire occurs
when the beam is on, record the following: the prescribed dosage and the backup
counter reading at the time of failure.
- Calculate the
remaining dosage by subtracting the backup counter reading from the prescribed
dose.
Treatment beam
failing to terminate
- Press the beam off
button and switch the enable/disable key to disable mode.
- If the beam is still
ON press the emergency button near the console and turn off the circuit
breaker.
- Remove the patient
from the treatment room.
- Record the backup
counter reading and calculate the remaining dose.
- Call a physicist or an
engineer to inspect the equipment
- The accelerator should
not be used for treatment until the problem has been solved.
Equipment
Malfunction
- Terminate beam and
record the backup counter reading at the time of failure.
- Remove the patient
from the treatment room.
- Call the engineer to
rectify the problem
- Calculate the remaining
dose and call the medical physicist to verify.
Accidental
Radiation Exposure
In case the linear
accelerator starts emitting higher radiation than the pre-set values, a medical
physicist or a Radiation Safety Officer should be informed immediately.
References
LSUHSC-Shreveport
Radiation Oncology. Linear Accelerator Emergency and Safety Procedures.
OzRadOnc. Emergency
Shut-Off. Retrieved from www.ozradonc.wikidot.com
on16/1/2019.
Home > Radiation Protection and Quality Assurance > Equipment Use & Quality Assurance > Quality Control Procedures > Emergency Switches
FREE Infographic What successful people
believe. What successful people do
Dictionary of Cancer Terms
Need help understanding a word? Here is an electronic resource that gives meaning to Cancer terms and their usage.
StrengthsFinder 2.0